DESIGNING AN IDEAL OPERATING ROOM COMPLEX : Indian Journal of Anaesthesia (2024)

Introduction

An operation theatre complex is the "heart" of any major surgical hospital. An operating theatre, operating room, surgery suite or a surgery centre is a room within a hospital within which surgical and other operations are carried out. Operating theatres were so-called in the United Kingdom because they traditionally consisted of semi-cir-cular amphitheatres to allow students to observe the medi-cal procedures.The Old Operating Theatre in London is one of the oldest, dating back to 1822 (Oxford English Dic-tionary and Wikipedia.com).

The patient is the centre point of a functioning OT complex. He / she is in isolation for varying times, away from his near and dear ones and is physically sick. Efforts are directed to maintain vital functions, prevent infections / promote healing with safety, comfort and economy.

The establishment and working of the operation the-atre (O.T.) needs specialised planning and execution and is not a simple civil engineering work. A "civil-mechanical-elec-trical-electronic- bio medical" combo effort driven and co-ordinated by the needs, preferences and safety of the medi-cal/ surgical team forms the basis for starting and maintain-ing an operation theatre. Anaesthesiologists, by virtue of their knowledge of the intricacies of physiology, physics and bio-medical aspects of medicine and constant proximity to the operation theatre should preferably be involved from the early stages of planning of operating theatres1.

On an average, operation theatres cater to 50%of the needs of total healthcare seekers. Anaesthesia-con-trolled time typically represents a small fraction (10-20%) of the total case time, the remainder (> 80%) being sur-geon and nurse-controlled2.

In the following pages, we will discuss the various considerations while planning an operation theatre com-plex. Functioning of an OT complex including scheduling, administration, staffing, sterilization techniques and data / audit management are out of purview of this topic.

Purpose of operation theatre complex

OT complexes are designed and built to carry out investigative, diagnostic, therapeutic and palliative proce-dures of varying degrees of invasiveness. Many such set ups are customized to the requirements based on size of hospital, patient turnover and may be speciality specific. The aim is to provide the maximum benefit for maximum number of patients arriving to the operation theatre. Both the present as well as future needs should be kept in mind while planning.

Different zones of OT complex

The location and flow of the patients, the staff and the materials form the three broad groups to be consid-ered during all stages of design3.

Four zones can be described in an O T complex4, based on varying degrees of cleanliness, in which the bac-teriological count progressively diminishes from the outer to the inner zones (operating area) and is maintained by a differential decreasing positive pressure ventilation gradi-ent from the inner zone to the outer zone.

  • (1) Protective zone: It includes
    • - Change rooms for all medical and paramedical staff with conveniences
    • - Transfer bay for patient, material & equipments
    • - Rooms for administrative staff
    • - Stores & records
    • - Pre & post-operative rooms
    • - I.C.U. and P.A.C.U.
    • - Sterile stores
  • (2) Clean zone: Connects protective zone to aseptic zone and has other areas also like
    • - Stores & cleaner room
    • - Equipment store room
    • - Maintenance workshop
    • - Kitchenette (pantry)
    • - Firefighting device room
    • - Emergency exits
    • - Service room for staff
    • - Close circuit TV control area
  • (3) Aseptic zone -Includes operation rooms (sterile)
  • (4) Disposal zone -Disposal areas from each OR & corridor lead to disposal zone

Sub areas (excluding OT place)

  • (1) Pre-operative check in area (reception)- This is important with respect to maintaining privacy, for changing from street clothes to gown and to provide lockers and lavatories for staff.
  • (2) Holding area- This area is planned for IV line in-sertion, preparation, catheter / gastric tube insertion, connection of monitors, & shall have O2 and suction lines. Facility for CPR should be available in this area.
  • (3) Induction room - (anaesthetic room). It should have all facilities as in OT, but there is controversy as to its need. One for each OT is required, ideally each is a duplicate of the other in each floor5. The anaesthetic room will provide a more tranquil atmo-sphere to the patient than the OT. It should provide space for anaesthetic trolleys and equipment and should be located with direct access to circulation corridors and ready access to the operating room. It will also allow cleaning, testing and storing of anaes-thesia equipment. It should contain work benches, sink(s). It should have sufficient power outlets and medical gas panels for testing of equipment.
  • (4) Post anaesthetic care units (PACU) -preferably adjacent to recovery room. These should contain a medication station, hand washing station, nurse sta-tion, storage space for stretchers, supplies and moni-tors / equipment and gas, suction outlets and ventila-tor. Additionally 80 sq ft (7.43 sq m) for each patient bed, clearance of 5 ft (1.5 m) between beds and 4 ft (1.22m) between patient bed sides and adjacent walls should be planned.
  • (5) Staff room -Men and women change dress from street cloth to OT attire; lockers and lavatory are essential; rest room TV, etc. are desirable.
  • (6) Sanitary facility for staff- One wash basin and one western closet (WC) should be provided for 8-10 persons. Showers and their number is a matter of local decision. Inclusion of toilet facilities in changing rooms is not acceptable; they should be located in an adjacent space4.
  • (7) The anaesthesia gas/cylinder manifold room/storage area- A definite area to be designated. It should be in a cool, clean room that is constructed of fire resistant materials. Conductive flooring must be present but is not required if non inflammable gases are stored. Adequate ventilation to allow leaking gases to escape, safety labels and separate places for empty and full cylinders to be allocated1.
  • (8) Offices - for staff nurse and anaesthesia staff- -The office should allow access to both unrestricted and semi-restricted areas as frequent communica-tion with public is needed.
  • (9) Rest rooms- Pleasant and quiet rest for staff should be arranged either as one large room for all grades of staff or as separate rooms; both have merits. Comfortable chairs, one writing table, a book case etc., may be arranged.4
  • (10) Laboratory - Small lab. with refrigerator for pa-thologist to be arranged.
  • (11) Seminar room- Since staff cannot leave an OT complex easily, it is better to have a seminar room within the OT complex. Intra-departmental discus-sions, teaching and training sessions for staff (with audio-visual aids) may be conducted here.
  • (12) Store room- This is designed to store large but less frequently used equipment in the OT. There should be storage space for special equipment after clean-ing.
  • (13) Theatre sterile supply unit (TSSU)- Within this area, following are desirable -
  1. Temperature between 180 -220 C, humidity of 40%--50% is the aim.
  2. Air conditioned with 10-12 air exchanges per hour
  3. Storage of sterile drapes, sponges, gloves, gowns and other items ready to use.
  4. Option to store in from one side and remove from other side.
  5. Proper inventory to prevent running out of stock.
  • (14) Scrub room- This is planned to be built within the restricted area. Elbow operated or infrared sensor operated taps / water source is ideal. It is essential to have non slippery flooring in this area.
  • Types of OT complexes

    There are three main categories of operating the- atres4:

    1. The single theatre suite with OT, scrub-up and gown-ing, anaesthesia room, trolley preparation, utility and exit bay plus staff change and limited ancillary ac-commodation.
    2. The twin theatre suite with facilities similar to 1, but with duplicated ancillary accommodation immediate to each OT, sometimes sharing a small post anaes-thesia recovery area.
    3. OT complexes of three or more OTs. with ancillary accommodation including post anaesthesia recovery, reception, porter's desk, sterile store and staff change.

    Principles to be taken into consideration while plan-ning an O.T. (physical / architecture):

    1. Location: Low rise buildings limited to two or three storeys high are preferred because of maximum advantage of natural light and ventilation as appro-priate can be derived4. The OT should be separate from general ′traffic′ and air movement of rest of the hospital, OT, surgical wards, intensive care units (ICU), accident and emergency department (A & E), Radiological department (X-Ray) should be closely related and access is also required to Sterilizing and disinfecting unit (SDU) and laboratory facilities. The location of the operation complex in a multi-storey building is planned on the first floor, connecting to surgical and other wards on the same floor. Adequate electric lift is planned for vertical movement from casualty on the ground floor and ENT, Orthopaedics, Ophthalmology & other wards on the floors above.
    2. Zone wise distribution of the area, so as to avoid criss-cross movements of men & machines
    3. Adequate & appropriate space allotted as per utility of the area
    4. Provision for emergency exit
    5. Provision for ventilation & temperature control, keep-ing in mind the need for laminar flow, HEPA filter air conditioner etc.
    6. Operation rooms:
    7. The number & size can be as per the requirement but recommended size is 6.5 m x 6.5m x 3.5 m. Glass windows can be planned on one side only.
      Doors: Main door to the OT complex has to be of adequate width (1.2 to 1.5 m). The doors of each OT should be spring loaded flap type, but sliding doors are pre-ferred as no air currents are generated. All fittings in OT should be flush type and made of steel.
      The surface / flooring must be slip resistant, strong & impervious with minimum joints (eg. mosaic with cop-per plates for antistatic effect) or jointless conductive tiles/ terrazzo, linoleum etc., The recommended minimum con-ductivity is 1m ohm and maximum 10m Ohms.
      Presently the need for antistatic flooring has dimin-ished as flammable anaesthetic agents are no longer in use.
      Walls- Laminated polyester or smooth paint provides seamless wall; tiles can break and epoxy paint can chip out. Collusion corners to be covered with steel or alu-minium plates, colour of paint should allow reflection of light and yet soothing to eyes. Light colour (light blue or green) washable paint will be ideal. A semi-matt wall sur-face reflects less light than a highly gloss finish and is less tiring to the eyes of OT team.
      Operation table: One operation table per OT
      Electric point: Adequate electric points on the wall (at < 1.5 m height from the floor) (discussed later)
      X-Ray illuminators: There should be X-ray film illu-minators preferably recessed into the wall.
      Scrub area: to be planned for atleast for 2-3 persons in each OT.
    8. There has to be a preparation room in clean zone
    9. Corridors not less than 2.85 m width for easy move-ment of men, stretcher & machines
    10. Separate corridors for uses other than going into OT.
    11. Rooms for different persons working in OT & for different purpose (it should be as per zone & size)
    12. Gas & suction (control, supply & emergency stock) for all OTs & areas where patients are retained. Oxygen, gas and suction pipe to be connected with central facility and standby local facility should also be available.
    13. Provision for adequate & continuous water supply:
    14. Besides normal supply of available water at the rate of 400 litres per bed per day, a separate reserve emergency over head tank should be provided for operation theatre. Elbow taps have to be 10 cm. above wash basins.
    15. Proper drainage system.
    16. Pre-operative area with reception with separate des-ignated area for paediatric patients is desirable.
    17. Adequate illumination with shadow less lamps of 70,000-120,000 Lumens intensity, for assessing pa-tient colour and tissue visibility (discussed under " Lighting").
    18. The safety in working place is essential, and fire ex-tinguishers have to be planned in appropriate zone.
    19. Provision for expansion of the OT complex should be borne in mind during planning stages itself.

    Recommendations on the number of OTs required

    It is observed that out of all surgical beds, of the hos-pital, 50%of patients are expected to undergo surgery. Thus for 100 beds, with average length of stay of 10 days for each patient, 10 operations per day can be performed.

    In general, multiuse OTs, instead of multiple OTs of-fer advantages of efficient man power utilization, economi-cal maintenance and better training of supporting staff.

    Thus, in a 300 bedded hospital (with 150 surgical beds), one OT complex with 3 OTs for General Surgery, Gynaecology, Orthopaedics/ENT, one for Endoscopy and one for Septic.

    OT will be required with 8 hours a day working dura-tion.

    Ventilation

    Ventilation should be on the principle that the direc-tion of air flow is from the operation theatre towards the main entrance4.There should be no interchange air move-ment between one OT and another. Efficient ventilation will control temperature and humidity in OT, dilute the con-tamination by micro-organisms and anaesthetic agents.

    There are two types of air conditioning systems: re--circulating and non re-circulating3.

    Non re-circulating systems heat / cool the air as de-sired and convey it into the operating room with ideally 20 air exchange per hour. Air is then exhausted to outside. Anaesthetic agents in the OT air are also automatically removed. These are thus ideal but are expensive.

    The circulating system takes some or all of the air, adjusts the temperature and circulates air back to the room.

    The broad recommendations include:

    • - 20-30 air exchanges / hour for re-circulated air
    • - Only upto 80%recirculation of air to prevent build up of anaesthetic and other gases
    • - Ultraclean laminar air flow - the filtered air deliv-ery must be 90%efficient in removing particles more than 0.5m m.
    • - Positive air pressure system in OT: It should en-sure a positive pressure of 5 cm H2 O from ceil-ing of OT downwards and outwards, to push out air from OT.
    • - Relative humidity of 40-60% to be maintained4
    • - Temperature between 200 -240 C. Temperature should not be adjusted for the comfort of OT per-sonnel but for the requirement of patient, espe-cially in pediatric, geriatric, burns, neonatal cases etc.

    Pendant services

    Two ceiling pendants for pipeline services should be designed; one for surgical team and one for anaesthetist4. Anaesthetic pendant should be retractable and have lim-ited lateral movement and provide a shelf for monitoring equipment. It should have oxygen, nitrous oxide, four bar pressure medical compressed air, medical vacuum, scav-enging terminal outlets and atleast four electric sockets.

    Piped gases in the OT

    1. Automatic / semi-automatic fail safe manifold room to be designed.
    2. Two outlets for O2 and suction and one for N2 O are a minimum in each OT.
    3. Pipeline supply system should be able to cut off from mainline if the problem occurs anywhere along the delivery hosing / tubing.

    Scavenging

    The method of scavenging should be decided during planning stage of OT3.US and International standards are available for scavenging but it is ideal to plan the type of system (active / passive) and number and location of scav-enging outlets beforehand.

    Electrical

    All electrical equipment in the OT need PROPER grounding

    In the past, isolated power systems were preferred when explosive agents were being used. They have the advantage of a transformer using grounded electricity and there is no risk to the patient or machines if a machine gets faulty.

    The grounded systems as used at homes offer pro-tection from macro shock but devices may lose power without warning. Life support systems, if in use could be disturbed.

    Following criteria are ideal with respect to electricity in OT complex:

    1. Use of circuit breakers / interrupters is desirable if there is an overload or ground fault.
    2. Power line of 220 Volts
    3. Suspended ceiling outlets should have locking plugs to avoid accidental disconnection.
    4. Insulation around ceiling electrical power sources should withstand frequent bendings and flexings. They should not develop cracks and should not damage wires. Wires inside rigid or retractable ceil-ing service column can help to some extent
    5. Wall outlets to be installed 1.5 m above ground (as already mentioned).
    6. Use of explosion proof plugs.
    7. Multiple outlets from different electrical line sources should be available.
    8. Electrical load calculation should be based on, equip-ments likely to be used and appropriate current car-rying capacity cords to be used.
    9. Emergency power: OT electrical networks need to be connected to the emergency generators with au-tomatic two way changeover facility.

    Lighting

    Some natural daylight is preferred by staff 4. Where possible, high level windows which give a visual appre-ciation of the ′outside world′ can be considered in the OT.

    1. General lighting: Colour corrected fluorescent lamps (recessed or surface ceiling mounted) to produce even illumination of atleast 500 Lux at working height, with minimal glare are preferred4. Means of dim-ming may be needed during endoscopies. To mini-mize eye fatigue, the ratio of intensity of general room lighting to that at the surgical site should not exceed 1:5, preferably 1:3. This contrast should be maintained in corridors and scrub areas, as well as in the room itself, so that the surgeon becomes accustomed to the light before entering the sterile field. Colour and hue of the lights also should be consistent.
    2. About 2000 Lux light is needed to asses the patients colour.
    3. White and glistening / shiny body tissues need less light than dark and dull tissues
    4. Operating area: Overhead light should be shadowless and give 25000-125000 Lux of light (50000 to100000 Lux at the centre and atleast 15000 Lux at the pe-riphery).
    5. About 10-12 inch of focus of light gives adequate illumination both at depth as well as surface of body.
    6. Lights should be freely movable both in horizontal and vertical ranges. Pendant systems are preferred. OT light should produce blue white colour of daylight at spectral energy range of 50000K (35000-67000- kelvin acceptable)
    7. Halogen lights produce less heat and hence pre-ferred. OT light should not produce more than 25000 mw/cm2 of radiant energy. Elimination of heat by dich-roic reflectors (cold mirrors) with heat absorbing re-flectors or filters should be available along with the luminaire4.
    8. An auxiliary light for a second surgical site is often beneficial
    9. UPS of adequate capacity to be installed after con-sidering OT light, anaesthesia machine, monitors, cautery etc until the backup generator takes over
    10. In endoscopic OTs, a reduced lighting is sometimes recommended. A grazing light over the floor can be helpful

    Anaesthesia equipment and monitoring needs

    At least one anaesthesiologist should be in the team involved in planning an OT. It is imperative that certain mandatory considerations with respect to the anaesthetic equipment and monitors be planned during the planning and design stage itself. Personal, practice and cost preferences may influence the plans.

    Communications

    Telephones, intercom and code warning signals are desirable inside the OT.3 One phone per OT and one exclu-sively for use of anaesthesia personnel is desirable. Inter-com to connect to control desk, pathology and other OTs as well as use of paging receivers (bleeps) is also ideal. A code signal, when activated, signals an emergency state such as cardiac arrest or need for immediate assistance.

    Catering

    Basic services such as preparation of beverages and some snacks, use of vending machines may be planned, augmented by provision of hot and cold meals from main hospital kitchen4.

    Cleaning

    The construction materials selected for the OT com-plex should aim to minimize maintenance and cleaning costs4.

    Data management

    Customised network connections should be put in place or a conduit should be planned. A well designed system can provide automated records, materials management, quality improvement and assessment, laboratory tracking, etc6. The Software for OT management are costly and hospitals are generally slow to adopt to changes. Customized OT soft-ware can be designed for individual needs.

    Operating theatre satellite pharmacy

    Access to the OT areas and outside should be pos-sible. It should have a laminar flow hood, a refrigerator, space for drug storage locked containers for controlled substances computer, desk area for paper work and phar-maceutical literature. Special kits for specific surgeries may also be arranged. The pharmacy may open for 1 to 24 hours based on need but it is desirable that an after hour system is planned.

    Statutory regulations

    The design and planning of an OT complex will need compliance with mandatory regulations related to local administration such as Municipal Corporation, Government, Pollution Control Board, Fire Safety Department, Water supply and Drainage department, etc.,

    Usual areas of deficiency in OTs (existing OTs)

    • 1) No reception area.
    • 2) No separate rooms for
    • - Surgeons
    • - Anaesthesiologist
    • - Jr. doctor
    • - OT attendants
    • 3) Not enough number of change rooms for differentclass of people.
    • 4) Inappropriate size & type of doors etc.,
    • 5) Lack of laminar flow & mandatory air exchange systems in OT.
    • 6) Lack of standard OT protocol.
    • 7) No separate Central Sterile Supply Department (CSSD)
    • 8) Waiting Area - Recovery
    • - Not well equipped
    • - Lack of basic amenities

    The authority for standardization

    Recommendations are available in various surgical, anaesthesia and nursing manuals with regard to the plan-ning and establishment of operation theatres / complexes. The hospital can get accredited by the Joint Commission on Accreditation of Healthcare Organizations (JCAHO)7, a professionally sponsored program that stimulates a high quality of patient care in health care facilities. There is also an accreditation option that is available for ambula-tory surgery centers (Accreditation Association for Am-bulatory Healthcare- AAAH). The department of health and social security (DHSS) in UK has publications con-taining information on planning for new health buildings and for upgrading existing buildings.

    Conclusion

    In the present era of evidence based medicine, it be-comes imperative to give maximum importance to plan-ning an Operation Theatre Complex. Within the limitations of finance and space, the best results can be obtained and anaesthesiologist with multiple roles inside the operation theatre complex, should be consulted in the process. Ef-forts should be made to conform to the standards laid down by local bodies and international agencies, as healthcare facilities in India are now catering to more and more inter-national clientele. However, new OTs and hospitals that are being established cannot be expected to fulfil all theo-retical requirements as new ideas are constantly being developed. By the time they are incorporated into build-ings, fresh ones take their place on the drawing board.

    1. Dorsch JA, Dorsch SE Operating room design and equipment selection. Understanding Anaesthesia Equipment. 19994th edition Williams and Wilkins:1015–16

    • Cited Here

    2. Barash P G, Cullen B F, Stoelting R K. Value based anesthesia practice, resource utilisation and operating room management Clinical Anaesthesia. 20014th edition Lippincott Williams and Wilkins:111–112

    • Cited Here

    3. Gupta S.K, Kant S, Chandrashekhar R. Operating unit - planning essentials and design Considerations Journal of Academy of Hospital Administration. 2005;17:01–12

    4. Bridgen RJ Ch.1.TheOperating department 2. Organisation and Management 3. Electricity & Electromedical Equipment 4.Static Electricity: Operating theatre technique, 5th edition: Churchill Livingstone. 1988 09,10,13,16-21,27-31,41,43-45,109

    • Cited Here

    5. Moyle JTB, Davey A, Ward C The anaesthetic room and recovery area.Ward's Anaesthetic Equipment. 19923rd Edition W B Saunder's:347–351

    • Cited Here

    6. Miller RD Operating room information systems. Miller's Anesthesia. 20056th Edition Elsevier -Churchill Livingstone:3131–32

    • Cited Here

    7. Joint Commission on Accreditation of Healthcare Organizations. (‘http://www.jointcommission.org/AccreditationPrograms/Hospitals/’)

    • Cited Here

    Keywords:

    Operating room; Operating theatre complex; Designing; Planning and organisation; Induction room; Accreditation.

    © 2007 Indian Journal of Anaesthesia | Published by Wolters Kluwer – Medknow
    DESIGNING AN IDEAL OPERATING ROOM COMPLEX : Indian Journal of Anaesthesia (2024)
    Top Articles
    Latest Posts
    Article information

    Author: Otha Schamberger

    Last Updated:

    Views: 5703

    Rating: 4.4 / 5 (55 voted)

    Reviews: 94% of readers found this page helpful

    Author information

    Name: Otha Schamberger

    Birthday: 1999-08-15

    Address: Suite 490 606 Hammes Ferry, Carterhaven, IL 62290

    Phone: +8557035444877

    Job: Forward IT Agent

    Hobby: Fishing, Flying, Jewelry making, Digital arts, Sand art, Parkour, tabletop games

    Introduction: My name is Otha Schamberger, I am a vast, good, healthy, cheerful, energetic, gorgeous, magnificent person who loves writing and wants to share my knowledge and understanding with you.